| Amino Acid Liquid Chromatography TR LC-AA98 / TR LC-AA98AT |
| Main configuration | Unit | Quantity | SEMI | AUTO |
| P98I Binary Gradient Pump | set | 1 | - | - |
| a. Operating pressure: 0~42MPa (6000 psi) maximum. b. Flow rate range: 0.001~10ml/min;c. Flow rate precision: ±1%; d. Flow rate accuracy: 25ml/368~387s(±3%); e. Compositional precision:0.5% absolute; f. Pressure solution: 0.1MPa; g. Over pressure protection: pump shuts off if pressure exceeds limit specified by user-42.0 MPa(6000 psi) maximum. |
| TR UV98I UV Detector | set | 1 | - | - |
| Wavelength range:190-700nm; Spectral bandwidth:5nm Wavelength accuracy:±2nm; Wavelength reproducibility:±0.2nm Flow cell volume:4.5μl; Time constants:0.05s, 0.5s, 5s |
| TR 7725i Injector | set | 1 | - | |
| TR Z98 automatic sampler | set | 1 | | - |
| Standard: 120×1.5 mL glassvials Number of repeati njection :1 to 99 times/sample During injection of 5 μL: CV 0.5% or less (n = 6) Cross contamination : ≤0.01%( According to specified wash needle procedure ) |
| C18, 5μm, Amino Acid Column | set | 1 | - | - |
| Guard Column | set | 1 | - | - |
| N2000 Work Station | set | 1 | - | - |
| T98 Column Oven | set | 1 | - | - |
| Temperature range: ambient ~150℃; Resolution: 1℃ steps; Precision: ±0.5℃ |
| Solvent Filter | set | 1 | - | - |
| 0.45μ Organic Filter membrane | piece | 100 | - | - |
| 0.45μ Water Filter membrane | piece | 100 | - | - |
| Organic Sample Filter | piece | 100 | - | - |
| Water Sample Filter | piece | 100 | - | - |
| Aspirator Vacuum Pump | set | 1 | - | - |
| Amino Acid Standard Sample | set | 1 | - | - |
| Amino Acid Derivatives Solvent | set | 1 | - | - |